Red Cycler 5C Instructions d'exploitation Page 312

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 377
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 311
07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 18-57
18. Troubleshooting
18.7 Reprime Patient Line Procedure
If the fluid level is not at or near the connector at the end of the patient line, follow
the steps below to reprime the patient line.
Steps to reprime the patient line Display screen
1. Press
STOP when the display screen alternates
between CONNECT YOURSELF and
CHECK PATIENT LINE.
alternates with
2. Press V until REPRIME PATIENT LINE appears.
3. Press
ENTER.
4. Verify that the patient line is properly primed:
a. Make sure fluid is present near the connector
at the end of the patient line.
The display screen alternates between
CONNECT YOURSELF and CHECK PATIENT LINE.
5. Repeat Steps 1 through 4 until the patient line is
primed.
CONNECT YOURSELF
CHECK PATIENT LINE
REPRIME PATIENT LINE
a
Vue de la page 311
1 2 ... 307 308 309 310 311 312 313 314 315 316 317 ... 376 377

Commentaires sur ces manuels

Pas de commentaire