9. Operating Instructions – Change Program
9-16 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
Last Fill
Volume
Last Fill Volume delivered at the end of the therapy and left in
the peritoneal cavity during the day. Also called “Wet Day.”
Minimum setting is 0 or 100 mL
Maximum setting is 3000 mL
Default setting is 0 mL
Setting increments:
10 in the 100 to 500 mL range
50 in the 500 to 1000 mL range
100 in the 1000 to 3000 mL range
Dextrose
(default setting)
– or –
The Last Fill Volume can be the same dextrose concentration
as the other Fills or it can be different.
This setting does not appear if LAST FILL VOL = 0.
This setting only appears if you use Last Fill.
Table 9-5. Hi-Dose CCPD Settings – Standard Fill Mode (Continued)
Setting Description
LAST FILL VOL: ML
DEXTROSE: SAME
DEXTROSE: DIFFERENT