Red Cycler 5C Instructions d'exploitation Page 285

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 377
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 284
18. Troubleshooting
18-30 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
18.3.15 Verify I-Drain
VERIFY IDRAIN
Message Display:
Cause 1: The I-DRAIN ALARM setting is OFF.
To Correct
Cause 1:
1. Press the power switch OFF and ON again.
2. Set the I-DRAIN ALARM to a value other than OFF. See
10.2.7, I-Drain Alarm, on page 10-9.
h NOTE: You have permanently changed the Initial Drain Alarm setting for
this therapy and all future therapy sessions.
Cause 2: The I-Drain Alarm setting is lower than expected.
To Correct
Cause 2:
1. Press
STOP and the setting blinks.
2. Press
U or V to temporarily change the minimum
I-Drain Alarm setting.
3. Press
ENTER to accept the value.
4. Press
GO to resume Initial Drain.
h NOTE: You have temporarily changed the Initial Drain Alarm setting for this
treatment session. To permanently change the I-Drain Alarm, see
10.2.7, I-Drain Alarm, on page 10-9.
VERIFY IDRAIN: OFF
Vue de la page 284
1 2 ... 280 281 282 283 284 285 286 287 288 289 290 ... 376 377

Commentaires sur ces manuels

Pas de commentaire